Ultrasound Doppler Systems For Examination of Peripferal And Transcranial Vessels |
Detection and quantification of intra—or extracrania stenoses/occlusions |
Price: £ ??? |
Applications |
The system is multifunctional diagnostic system, which is intended for · CW Doppler examinations of the peripheral and extracranial vessels · PW Doppler examinations of the intracranial vessels · Culation examinations of uteroplacental blood flow circulation
|
. |
· long time monitoring of transcranial vessels with automatic emboli detection · monitoring ECG and Doppler signals simultaneously · display trends of all Doppler parameters and indicies
|
Multiwindow screen layout |
Report editor |
Report editor allows to prepare quick report on the base of the report templates, which include: · spectra of blood flow · hospital name · ID of patient · index table · date and time of examination ID of physician |
Examinations |
The following items are calculated automatically during examination: · Blood flow curves · Sistolic and diastolic marks · Index tables for every heart cycle · pulsation index for the brain structure pulsations Spectrum and/or curves can be displayed in the spectral window. One of the two modes can be used to display spectrum from the data base: o Cineloop with sound tracking o Manual sliding under spectral buffer
|
A manual on Doppler examination of cerebral vessels. Software package includes build-in medical help (in hypertext format). This medical help contains a series of normal and pathologic Doppler patterns which correspond of various states of the cerebral arterial bed. Each Doppler pattern is followed by description, diagnostic and therapy steps. Except pattern descriptions, medical help contains: · Information about anatomy of the artery · Method of it's location · Protocol of the Doppler examination
|
Up to 10 windows can be displayed simultaneously on the screen. Flexible screen layout (from 1 to 10 windows) simplifies: · carrying out examinations, especially in complex cases · comparison spectra of left/right side vessels · Comparison of the current examaination spectra and spectra from database |
TEMPORARY SIGMATRON SITE
|
Additional software package is intended for long monitoring of transcranial vessels together with monitoring of ECG signal. Some features of software package are below: · Automatic emboli detection with generating of histogram · Trend window for all indices and parameters of blood flow · Real time marking of different events during monitoring · Generating event table for further data analysis The one of the three data storage format can be seleted to save monitoring data (brief, middle, complete). |
Additional software package |
A manual on Doppler examination of cerebral vessels |
Accessories |
Doppler unit |
|
Doppler probes: |
2, 4, 8 MHz |
Ultrasound power: |
2 MHz < 1.0 mW/mm2 (PW) |
|
4 and 8 MHz < 0.4 mW/mm2 (CW) |
Frequency rangeт (adjustable): |
100 Hz - 20 kHz |
Wall fiters : |
100, 200, 400, 800 Hz |
Adjustable PRF : |
5 - 16 kHz |
Depth: |
30-130 mm (1 mm step) |
Sample volume : |
3-20 mm (1 mm step) |
Spectrum analyzer : |
128 /256 point FFT |
Visualization: |
512 spectral lines / 256 frequencies, 32 color / grey levels |
Adjustable smoothing and compression of spectral data |
|
Real-time index calculating |
(RI, PI, ISD, STI, HR) |
Real-time curve calculating and drawing |
(Vs, Vd, Vave) |